SE-2003&SE-2012 Holter System Recorder User Manual
- 18 -
3.4.1.3 7 Electrodes 3 Channels
Seven-electrode-three-channel lead wires are utilized by the recorder to create a 3-channel ECG
signal. Different placement positions of 7 electrodes will form different lead modes, and it is
recommended to place electrodes as below:
IEC AHA Electrode Placement
Green
(CH1+)
Red (CH1+) Left Anterior Axillary line 6
th
rib.
Red (CH1-)
White
(CH1-)
Right Manubrial border of the Sternum.
White
(CH2+)
Brown
(CH2+)
Approximately 1 inch right of Xiphoid Process on the rib.
Yellow
(CH2-)
Black
(CH2-)
Left Manubrial border of the Sternum.
Black (RL) Green (RL) Lower right rib margin over bone.
Orange
(CH3+)
Orange
(CH3+)
Left Mid-Clavicular line 6
th
rib.
Blue
(CH3-)
Blue (CH3-) Center of the Manubrium.