SE-2003&SE-2012 Holter System Recorder User Manual
- 31 -
Press OK to enter next step. Press Reset to reset the patient information.
5.2.3 ECG Preview
The process will be skipped if the setting of ECG preview is off under Advance Setting menu.
The upper part displays one-channel ECG wave, while the bottom displays name of the lead
(channel) ( I, II, III, aVR, aVF, aVL, V1, V2, V3, V4, V5, V6 or CH1, CH2, CH3 in 3-channel
mode). Press left-up key or right-down key to change ECG lead (channel), and press confirmation
key to enter next process.
NOTE: The time axis (paper speed) in waveform display mode is around 1.5cm/s or
3cm/s, which is slightly different from regular ECG with paper speed of 2.5cm/s or
5cm/s. Actually the main purpose is to visually assess signal quality and basic
form, but not to be basis of quantitative measurement.
5.2.4 Confirmation
Shift cursor between OK and Exit by pressing left-up key or right-down key. Press Exit to return
to main menu or press OK to start recording.
ID: 2345604560
Name: Mary
OK Reset
Start Recording ECG?
OK Exit