EasyManua.ls Logo

EDAN SE-2003 - Technical Specifications; Safety Compliance and Standards; Environmental Operating Conditions

EDAN SE-2003
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SE-2003&SE-2012 Holter System Recorder User Manual
- 41 -
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC 60601-1:1988+A1:1991+A2:1995
EN 60601-1:1990+A1:1993+A2:1995
IEC60601-1-2:2001+A1:2004
EN 60601-1-2:2001+A1:2006
IEC/EN 60601-2-47
Anti-electric-shock type: Internal power supply
Anti-electric-shock degree: Type CF
Degree of protection against
harmful ingress of water:
Ordinary equipment (Sealed equipment without liquid
proof)
Disinfection/sterilization
method:
Refer to the user manual for details
Degree of safety of
application in the presence of
flammable gas:
Equipment not suitable for use in the presence of
flammable gas
Working mode: Continuous operation
EMC: Group 1, Class B
Patient
Leakage
Current:
NC <10μA (AC) / <10μA (DC)
SFC <50μA (AC) / <50μA (DC)
Patient
Auxiliary
Current:
NC <10μA (AC) / <10μA (DC)
SFC <50μA (AC) / <50μA (DC)
A1.2 Environment Specifications
Transport & Storage Working
Temperature:
-20 (-4 ºF) ~ +55
(+131 ºF)
+5 ºC (+41 ºF) ~ +40 ºC
(+104 ºF)
Relative Humidity:
25%~93%
Non-Condensing
25%~80%
Non-Condensing

Table of Contents