SE-2003&SE-2012 Holter System Recorder User Manual
- 22 -
4.3.2 Patient ID Writing
E-label registration detection can be set to on or off. Default value is the value set last time.
When set to on, SD card will be checked whether E-label is registered during start-up process. If
not, the recorder will enter Patient ID writing process and you need to input Patient ID through
keyboard.
When set to off, the recorder will skip Patient ID writing process, which will not affect normal
E-label registration process.
If more than one patient is tested at the same time, it is recommended to set the item as on.
If only one patient is tested and registration process is found to be troublesome, the item can be
set to off so as to simplify the process.
4.3.3 ECG Display
ECG display during recording process can be set to on or off. Default value is the value set last
time.
Set the value as on and enter ECG display mode by pressing Left-up key and confirmation key
during recording process and observe waveform.
It is unable to enter ECG display mode by pressing Left-up key and confirmation key during
recording process if the value is set to off. This helps to prevent unnecessary influence to patients.
Whether entering ECG display mode or not, recording signals will not be influenced.
4.3.4 LCD Contrast
LCD contrast can be adjusted in the range of 10%~100, with 10% of grade difference.
Default value is the value set last time.
The item is adjusted according to light environment and user habits. There is no need to modify it
under normal situations.
4.3.5 Data Protection
Data protection is used to detect if the recorded data has been analyzed. If it is set to on, the
recorded data has to be analyzed by analysis software at least once before starting the next
recording. It can also be set to off. The last configuration will be the default.