6.2 Maintenance plan
The plan shown in Table: Maintenance plan on page 34 details the roune maintenance
operaons necessary to maintain the pump in normal use. Instrucons for each
operaon are given in the secon shown.
More frequent maintenance may be required if the pump is used to process corrosive or
abrasive gases and vapours, in these circumstances, we recommend you to replace the
pump seals every year. If necessary, adjust the maintenance plan according to
experience.
When carrying out maintenance on the pump, use our spares and maintenance kits,
these contain all the components necessary to complete maintenance operaons
successfully. The item numbers of the spares and kits are given in Spares.
Examine the condion of any external accessory, lters or traps (if ed) when carrying
out maintenance on the pump. Refer to the instrucons supplied with these accessories
for the necessary maintenance procedures.
Table 10
Maintenance plan
Operaon Frequency Refer to secon
Check the oil level As required; at least monthly Check the oil level on page 34
Replace the oil 3 to 6 months depending on
applicaon
Replace the oil on page 35
Inspect and clean the inlet lter Every oil change Inspect and clean the inlet-lter
on page 35
Clean or replace the gas-ballast
lter
Yearly Clean the gas-ballast lter on
page 36
Clean the motor fan-cover and
enclosure
Yearly Clean the motor fan-cover and
enclosure on page 37
Clean and overhaul the pump and
replace the blades as necessary
3 to 4 years depending on
applicaon
Clean and overhaul the pump on
page 37
Replace the run capacitor Every 4 years Replace the run capacitor and test
the motor on page 37
Test the motor condion Every 4 years Replace the run capacitor and test
the motor on page 37
Fit new blades Every 30000 hours of
operaon
Fit new blades on page 38
6.3 Check the oil level
Note:
1. If required, it is possible to check the oil
level while the pump is operang, however
the pump must be switched o. Isolate the pump and other components in the pumping
system from the electrical supply before pouring oil into the pump.
2. Do not mix hydrocarbon lubricants with PFPE or vice versa. If the oil is mixed, drain
and rell with clean oil as described in Replace the oil on page 35.
Refer to Figure: General view of the pump on page 11 for the items in brackets.
Page 34
A37310880_V - Maintenance

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards E2M30 and is the answer not in the manual?

Edwards E2M30 Specifications

General IconGeneral
BrandEdwards
ModelE2M30
CategoryWater Pump
LanguageEnglish

Related product manuals