8. Storage
CAUTION:
Observe the storage temperature limits stated in Technical data on page 12. Storage
below - 30 °C will permanently damage the pump seals.
Note:
If you will store a new pump in condions of high humidity, remove the pump from its
cardboard packaging box, dispose of the box (refer to Disposal on page 42).
Use the following procedure to store the pump:
1.
Ensure that the pump has been shutdown as described in Operaon.
2.
Isolate the pump from the electrical supply.
3.
Purge your vacuum system and the pump with dry nitrogen and disconnect the
pump from your vacuum system.
4. Replace the oil as described in Replace the oil on page 35.
5. Place and secure protecve covers over the inlet and outletports.
6.
Store the pump in cool, dry condions unl required for use.
7.
When required, prepare and install the pump as described in Installaon. If the
pump has been stored for more than a year, before you install the pump you must
clean and overhaul it as described in the instrucon supplied with the clean and
overhaul kit.
Page 41
A37310880_V - Storage