© Edwards Limited 2010. All rights reserved. Page 19
Edwards and the Edwards logo are trademarks of Edwards Limited.
Installation
A541-11-880 Issue A
3.7.2 Gas purges
Gas purge is provided for the sole purpose of enabling the pumping of inert light gases such as helium.
3.8 Connect the electrical supply to the dry pumping system
WARNING
The iXL120 and iXL120N pumps are not suitable for use with hazardous production materials. Any
hazardous production materials must be diluted to a safe level before entering the pump.
WARNING
Ensure that the electrical installation of the pumping system conforms with your local and national
safety requirements. It must be connected to a suitably fused and protected electrical supply and
a suitable Earth (ground) point.
WARNING
This equipment is suitable for Installation Category II as defined in IEC 60664-1. The dry pumping
system must be connected to an isolator that disconnects all current carrying conductors and can
be locked in the off position. The isolator must be in close proximity to the equipment, within easy
reach of the operator and identified as the disconnect device for the equipment.
WARNING
The equipment must be permanently Earthed. A second protective Earth (ground) conductor, with
a cross-sectional area at least equal to the phase conductor size, must be connected to the
protective Earth (ground) stud.
WARNING
Ensure that the system and your electrical supply cable are suitably protected against Earth
(ground) faults and that the Earth (ground) conductor of the electrical supply cable is longer than
the phase conductors in the connector.
WARNING
All connections to the interface control must be double insulated or have equivalent protection.
Do not connect voltages greater than 30 V a.c. or 60 V d.c. to the control/interface connections.
If you do, the interface control will not provide protection against electric shock.
WARNING
The power wiring must be protected, refer to Table 8 for requirements.
WARNING
The main power connector is not approved for connection and disconnection under load.