Installation
A541-11-880 Issue A
Note: Plug/Socket designation refers to the pins of the connector.
3.10.1 Inputs
Notes: 1. Volt free contacts sensed at 24 V / 5 mA, these inputs should not be linked together.
2. Voltage inputs accept AC or DC voltage (except SPI pins 15 and 16), between 15 V and 24 V selects the
input as active, below 5 V selects inactive.
3. To ensure correct pump operation, at least 10 seconds must be allowed between successive operations
of the input signals.
3.11 Connect the cooling-water hoses
Water only flows if the pump is running or for about 10 seconds when the pump is first energised. You may
damage your water circulation system if it cannot tolerate a no flow condition.
Note: For optimum water-cooling, ensure that your cooling-water supply complies with Section 2 and that the dry
pump is not connected in series with any other equipment.
Connect the cooling-water supply as follows:
1. Connect your supply and return hoses to the pump.
2. Turn on the cooling-water supply.
3. Energise the system to ensure water flows through the pump.
Table 13 - Tool connectors and mating tool connector kits
Identification Part Number Tool Side Connector Tool Connector Kit
SPI D37360310 CPC 16/17 Socket D37420801
MCM D37360320 CPC 16/17 Socket D37422802
TEL D37360330 25w D Type Socket D37420802
E73 D37360340 15w D Type Plug D37420803
LAM D37360350 25w D Type Plug D37422801
C3 D37360360 9w D Type Socket D37421803
HIT D37360370 37w D Type Socket D37421804
Table 14 - Tool input signals to the pump system
Function Signal Type Pins Operation
Pump On/Off Voltage input SPI - 1 and 2 Pump runs when voltage
applied
MCM - 1 and 2
LAM - 10 and 23
C3 - 2 and 6
Volt-free contact input MCM - 5 and 6 Pump runs when contact
closed
TEL - 1 and 14
E73 - 1 and 9
HIT - 7 and 26 (or HIT - 1
and 20)