Note:
For
opmum water cooling, make sure that the cooling water supply meets the
specicaon given in Cooling water data on page 21. Make sure water supplies are
connected in parallel. Refer to Figure: Controls/connectors on the rear of the system. We
recommend use of quick connectors to decrease the risk of water spills during connecon
and disconnecon. Some pump variants are supplied with quick connector mang
halves. The quick connector mang halves are also available as an accessory kit. Refer to
Accessories on page 93.
For minimum water consumpon, control the cooling water ow to the system.
Seng the cooling water ow rate - Method 1
Before you start, make sure that the electrical power supply to the system is set to o.
To connect the cooling water supply:
1. Use a 3/8 inch BSP male pipe ng to aach the female quick-release connector
to the cooling water supply hose.
2. Use a 3/8 inch BSP female pipe ng to aach the male quick-release connector
to the cooling water return hose.
3. Remove the dust caps from the cooling water inlet and outlet.
4. Connect the water return hose to the cooling water outlet. Refer to Figure:
Controls/
connectors on the rear of the system
5. Install a water ow meter into the water supply line close to the pump and then
connect the water supply hose to the cooling water inlet. Refer to Figure: Controls/
connectors on the rear of the system
6. Turn on the cooling water supply.
7. Switch on the electrical power to the system. The water valves in the pump cooling
system will automacally open shortly aer power up (in 1 minute) for a period of
20 seconds.
8. While the main pump water valves are open, adjust the water ow r
ate so that it
meets the requirements given in Table: Water cooling system data.
Note:
All the water valves will close again aer 20 sec
onds and the ow rate displayed by
the water ow meter will decrease. If necessary, cycle the power to the pump to re-
open the valves for a further 20 seconds to connue seng the water ow rate.
9. Remove the water ow meter aer the water ow rate is set.
10. Do the inspecon and leak check of the water hoses, pipelines and connecons.
11. Turn o the water supply.
Seng the cooling water ow rate - Method 2
Use the following procedure to connect the cooling water supply and make sure that the
system is receiving the correct water ow rate. Before you start, make sure that the
electrical power supply to the system is switched o:
10/2020 - ©Edwards Limited
Page 46M56635880_D
M56635880_D - Inst
allaon

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards iXM Series and is the answer not in the manual?

Edwards iXM Series Specifications

General IconGeneral
BrandEdwards
ModeliXM Series
CategoryWater Pump
LanguageEnglish

Related product manuals