3.2 Operaon and s
torage environment
Table 6
Operaon and storage environment
Range Data
Ambient operang temperature range 5 °C to 35 °C
Ambient operang humidity range 10 to 90% RH (noncondensing)
Maximum operang altude 3000 m
Ambient storage temperature range -30 °C to 70 °C
3.3 Vent gas specicaon and vent control data
The pump can be vented to atmosphere, but the high relave humidity of the air can
increase the subsequent pumpdown me. To decrease the pumpdown mes, vent
with dry and clean gases. Refer to Vent opons, vent valve connecon and control on
page 57 for the vent opons and the vent valve connecon. Refer to Controlled venng
opons on page 64 for conguring the venng opons.
Table 7
Vent gas specicaon and vent control
Vent gas specicaon and control Reference data
Vent gas Dry air, nitrogen, argon or other inert
gases
Maximum dew point at an atmospheric
pressure
-22 °C
Maximum size of parculates 1 µm
Maximum concentraon of oil 0.1 parts per million
Recommended me for the rotaonal speed
to reach 50%
> 15 seconds
Maximum permied rate of increase of
pressure
Refer to Figure: Maximum allowed rate
of increase in pressure during venng:
pressure against me (pump inially at
full speed).
04/2022 - ©Edwards Limited
Page 27B80000880_G
B80000880_G - Technical data