Cooling water specicaon R
eference data
pH value 6.0 to 8.0
Maximum calcium carbonate concentra-
on
75 parts per million
Maximum chloride concentraon 100 parts per million
Minimum oxygen concentraon 4 parts per million
Minimum cooling water ow rate (at
15 °C)
15 l hr
-1
Water temperature 10 to 20 °C
Maximum water pressure
5 bar (gauge), 73.5 psig, 6 x 10
5
Pa
Ma
terials exposed to the cooling water Nickel plated brass
3.6 Electrical data
The pumps can be operated by the customers system or by the manufacturer's TIC Turbo
Instrument Controller or TIC Turbo Controller.
If using the customer system, the size of the power supply depends on the applicaon.
The rate of increase in the pump speed is dependent on the power limit seng. The
power limit seng overrides the necessary power supply seng. If the serial
communicaons or an access to the manufacturer's TIC is available, the power limit
seng of the pump can be selected. Refer to Table: Logic interface technical data for
the maximum power limit sengs for the pumps. If the rapid cycling of the pump is
necessary for an applicaon, you can get the faster rate of increase in the pump speed if
the power supply supplies higher current, up to a maximum in accordance with Table:
Logic interface technical data.
If the power limit seng is not adjustable, use a power supply capable to supply
sucient current to meet manufacturer's factory default power limit seng shown in
Table: Power limit seng.
The manufacturer's TIC has variants. If the pump is operated using the
manufacturer's TIC, contact us to select the applicable variant for the applicaon.
3.7 Pumping media
WARNING: DANGEROUS GASES
Risk of asphyxiaon. R
elease the dangerous gases and gas mixtures safely. Do not
expose people to the gases. If hazardous gases or vapour must pump, obey the safety
recommendaons of the supplier of the gas or vapour.
WARNING: PYROPHORIC OR EXPLOSIVE GASES
Risk of explosion. Do not use the nEXT pump t
o pump pyrophoric or explosive gas
mixtures as the pump is not applicable for this purpose. The pump and its
connecons are not designed to contain an explosion.
04/2022 - ©Edwards Limited
Page 29B80000880_G
B80000880_G - Technical data