EickTron120‐160
EN‐15
MA510a_EN
Instruction’sManual Manualed’Istruzioni
DEUTSCH
PORTUGUES FRANCOIS ESPANOL ENGLISH ITALIANO
SAFETY
WARNING:Electrosurgerycanbedangerous.Carelessuseofanyelement
in the electrosurgical system may subject the patient to a serious burn.
Read and understand all warnings, precautions, and directions for use
beforeattempttouse anyactiveelectrode.Neithermanufacturer,canbe
consideredresponsible forpersonal,materialorconsequentialinjur
y,loss
or damage that results from improper use of the equipment and
accessories.
Theaccessoriessuppliedwiththeunithavecharacteristicscompatiblewith
this supplied unit, they could be incompatible with others electrosurgical
units; the user must check, before connecting other accessories to this
unit,thatthey have characteristicsof insul
ation compatiblewiththoseof
thisunitandutilizedfunction(seeTechnicalCharacteristics).
It is recommended to inspect the integrity of the packagingof the sterile
products.
General
Thefollowingprecautionsreducetheriskofaccidentalburnings
The whole surface of the patient plate must be placed on a well
vascularized muscle as next as possible to surgical area. Avoid
connectingthepatientplatetobonyprotrusions,prosthesis,cicatricial
tissues,andpartsofthebodysubjectedtoliquidaccumulationorthat
present subcutaneous adipose tissue. The part of the body must be
withouthair,dryandclean.Donotusealcoholtocleantheskin.Unless
forveterinaryuse,theuseofgelatinoidssubstancesfortheelectrodes
isnotadvised.
Byusingthedisposableneutralelectrodesrespectthedateofexpire.
By usingthereusableelectrodesascertainthatthefixingsystemsgive
warrantyofstability.
When you apply the neutral electrode avoid the transversal course
andprefertheverticalordiagonalcourse,inparticularifasplitneutral
electrode is used. That to allow a uniform distribution of the current
onthesurfaceoftheneu
tralelectrodeandreducetheriskofburnto
thepatient.