Repairs
High frequency cables and electrode holder handle cannot be repaired.
Alwayssubstituteadamagedpartwithanewone
.
FusesSubstitution
Beforesubstitutingthefuse,disconnecttheunitfromthemainssystem
Only use fuse of the kind 5x20; they must have those characteristics:
T3.15A (slow) (230Vac mains voltage), T6.3A (115Vac mains voltage),
proceedasfollows:
(AB)Extractthefuseholderdrawerfromthepowermodule.
(C)Insertthefusesmakingre
ferencetothefollowingchart:
MainsVoltage 110120V DelayedFuseT6,3AL,250V/5x20mm
MainsVoltage 220240V DelayedFuseT3,15AL,250V/5x20mm
(D)Extractandrotatethedetachablepartinwaytoreadthe
correctvoltageinthe(E)windowreinsertthefuseholderinthe
module.
Fi 1
CheckingoftheEquipmentBeforeEachUse
Each time the use of the electrosurgicalequipment is planned a checkof
the most important safety aspects has to be implemented considering at
leastthefollowing
:
Checktheintegrityofcords,connections,wiresbreakage,etc.
Assurethatalltheelectricalequipmentisproperlygrounded.
Assurethatalltheaccessoriesthatshouldbeusedareavailableand
sterilized.