EN‐32
MA510a_EN
Instruction’sManual
PreselectionoftheDeliverableCurrent
The deliverable current for the surgical operations can have preselected
throughpushbuttonfor
:
CutCurrent(CUT)
The best current for the cut is the pure sinusoidal wave without
modulationthatmeanswithdutycy cle100%.Suchcurrent,proper
forcutwithoutcoagulation.
CoagulatedCutCurrent(BLEND)
The coagulatedcut current (BLEND) it is suited for coagulated cut
whenadeepcoagulationtogetherthecutisdesired.Thiscurrentis
madebysinecurrentesuitedforthecutassociatedtolowvoltage
current suited for coagulation(soft coag). With this blending, a
current suited for cut coagulated in abs
ence of eschar and
carbonizationisobtained,particularlysuitableforendoscopi csurgery.
SuperficialCoagulationCurrent(FORCEDCOAG)
Themodulatedcurrent(FORCEDCOAG)itischaracterizedbygood
property of surface coagulation behaving at the time it probable
production of eschar and partial carbonization of the tissue. The
advantage of this type of coagulation resides in the rapidity with
whichtheeffectisgotte n.