OperationMode
SwitchOn
Whenswitchedontheelectrosurgicalunitperformsautomaticallyatestto
establish the correct operation of itself and of the connected accessories
as well. In case anomaly is found an alphanumeric message it is shown
codedaccordingtothechartcodesbroughtinthechapterMAINTENANCE.
Thistestlastsabout10seco
nds.Atthe endofthecontroltheequipment
restoreslastuseoperationalconditions.
NeutralElectrode’sCircuit
The neutral electrode is continually watched by a special circuit that
prevents,danger of burnsto thepatientduethe loss of contact between
theneutralelectrode and thepatientskin,ifsplitelectrodeisused.If the
impedancevalueofthe patientcircuitismore than 200 ohm the value is
notaccepte
d,inthiscasetheOCindicatorlightflashing,andiftheoutput
circuitisactivatednopoweroutandthereisasoundalarm.
Inorder to reducetheacoustic pollution, thesound alarmis present only
whenpressedthefoot‐switch.
Ifasingleplateelectrodesusewatchedonlytheco
nnectionoftheneutral
electrodeplatetotheunit.