Do not reduceor disable theaudiblesignalwarning the activation of
thegenerator.Afunctioningactivationsignalcanminimizeorprevent
patientorstaffinjuryintheeventofaccidentalactivation.
Avoid verifying the functioning of the unit by shorting the active
electrodewiththereferenceoneortheacti
veelectrodewithmetallic
parts.
Ifnecessaryuseasmokeplumeextractionsystem.
SafetyforthePatient
During the HF electrosurgical operations the patient is a conductor of
electricalvoltage against earthpotential.So if there is acontactbetween
patient and electrical conductive objects (metal, wet clothes,etc..), in the
contact’s point could be electrical current that causes thermical necrosis.
Soitisrecommendedtoinspecttheequipmentanditsa
ccessoriesbefore
usingandtorespectallsafetyrules.