EickTron120‐160
EN‐21
MA510a_EN
Instruction’sManual Manualed’Istruzioni
DEUTSCH
PORTUGUES FRANCOIS ESPANOL ENGLISH ITALIANO
Before to apply the neutral electrode, clean and eliminate any external
substancesfromitssurface.
Do not apply the neutral electrode on cicatrix, bony protrusion or near
prosthesisormonitoringelectrodes.Butapplyitonsprinkledtissues,such
as muscles and near the operative site. If you use a disposable neutral
electrode res
pect the date of use, if you use a not disposable neutral
electrodemakesurethatthefixingsystemsguaranteestability.
Itisveryimportantthattheneutralelectrodeisfirmlyappliedonitsentire
surface to avoid burns. When the neutral electrode is partially taken off
from the patient, th
e current density on the remaining applied part is
higher.Becausethedensityofthecurrentflowundertheneutralelectrode
isnotuniform,itverifiesanotuniformheating,especiallyneartheborders
oftheneutralelectrode.