Before to apply the neutral electrode, clean and eliminate any external
substancesfromitssurface.
Do not apply the neutral electrode on cicatrix, bony protrusion or near
prosthesisormonitoringelectrodes.Butapplyitonsprinkledtissues,such
as muscles and near the operative site. If you use a disposable neutral
electrode res
pect the date of use, if you use a not disposable neutral
electrodemakesurethatthefixingsystemsguaranteestability.
Itisveryimportantthattheneutralelectrodeisfirmlyappliedonitsentire
surface to avoid burns. When the neutral electrode is partially taken off
from the patient, th
e current density on the remaining applied part is
higher.Becausethedensityofthecurrentflowundertheneutralelectrode
isnotuniform,itverifiesanotuniformheating,especiallyneartheborders
oftheneutralelectrode.