Monopolartypicalconfiguration
UseforMONOPOLAR:
Holderhandle with two pushbuttons
without foot switch: press the yellow
pushbutton on the holderhandle to
deliver the cutting current (the choice
between CUT or BLEND must be done
pressing the correspondent pushbutton on the unit); or the blue
pushbuttonontheholderhandletodelivercoagulatingcurrent(th
echoice
between FORCED COAG, SOFT COAG or BIPOLAR must be done pressing
thecorrespondentpushbuttonontheunit).
Holderhandlewithtwopushbuttonsand
a single foot switch: choose the cutting
current CUT or BLEND and the
coagulation current FORCED COAG, SOFT
COAG or BIPOLAR. Preset through the
yellow push
button on the holder handle,
the function for the cut that appears on
the unit or, through the blue pushbutton on the holder handle, the
functionforthecoagulationthatappearsontheunit.Thecurrentdelivery
takesplacethroughthefootswitch.