USER MANUAL MAGIC 500 PLUS ULTRASOUND UNIT
T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
4. OPERATION CONDITION
4.1 Power Supply
Adapter ratings: 100 240V~, 1,2 0,6A, 50 60Hz
Output of adapter: DC12,8V, 3,0A
Main device rating: DC12V, 3,0A
Internal supply voltage: DC11,1V ± 10 %
4.2 Operation Environment
Ambient temperature: 10 °C 40 °C
Relative humidity: 30 % 75 % (without condensation)
Atmospheric pressure: 700hPa 1.060hPa
Altitude: < 2.000 m
Overvoltage: Overvoltage Category II
Pollution degree: 2
4.3 Storage and Transportation
Ambient temperature: -20 °C 55 °C
Relative humidity: 30 % 93 % (without condensation)
Atmospheric pressure: 700hPa 1.060hPa
Danger!
Do not use this equipment where fl ammable gas (such as anesthetic gas, oxygen or hydrogen) or ammable liquid
(such as alcohol) are present. Failure to do so may result in explosion.
Warning!
Avoid using this equipment with high-frequency electric knifes, high-frequency therapy equipment or defi brillators
and other electronic devices, or an electric shock may occur to the patient.
Attention!
The mains voltage varies with different countries or regions.
Using radio transmitting equipment nearby the system may interfere with the normal operation of the system.
It is prohibited to carry or use devices that can generate radio waves within the room this system is installed,
such as cell phones, radio transceivers and wireless remote control toys.
System should not be used in following environments:
1. Moist
2. Rain
3. Thunderstorm weather
4. No ventilation
5. Close to heat source (e. g. heaters, microwave ovens, ovens, water heaters, etc.)
6. Direct sunlight
7. Dramatic temperature change
8. Poisonous gas
9. Corrosive gas
10. Strong shock
11. Strong electromagnetic eld (e. g. MRI)