USER MANUAL MAGIC 500 PLUS ULTRASOUND UNIT
T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
11. SYSTEM MAINTENANCE
The system maintenance should be performed by the user. Users shall be in full charge of maintenance and operation of the
system after purchasing the product.
Warning!
The following parts of the equipment are not serviced or maintained while in use.
11.1 Maintenance by Users
11.1.1 System Cleaning and Disinfection
Warning!
Turn off the instrument and pull out the power supply wire before cleaning. It may cause electric shock when
cleaning the system while the power is on.
The waterproof grade of the equipment is IPX4 (no adverse effect on splashing water in all directions). Do not
spill water or liquid into the device during cleaning or maintenance. Failure to do so may cause malfunction.
Do not place the ultrasonic probe connector into water or disinfection, as it may cause electric shock or
malfunction of probe.
Attention!
1. To prevent possible infection, it is advisable to wear sterilized gloves when cleaning or disinfecting the
ultrasonic probe.
2. In the process of cleaning and disinfection, avoid probe overheat (exceeding 60 °C) as it may be damaged or
deformed under excessive heat.
1. Clean the probe
1. Must wear sterilized gloves to prevent possible infection.
2. Rinse the probe with water or soapy water to remove all contaminants, or use a soft urethane sponge to wipe the probe.
Do not use brushes as this may damage the probe.
3. After nishing the rinsing, use a sterilized cloth or gauze to wipe the water on the surface of probe. Do not dry the probe
by heating it.
2. High-level disinfection
Please follow the disinfection method provided in this user manual for disinfection.
1. Before disinfection, wear sterilized gloves to prevent possible infection.
2. You must clean the probe before disinfection. Recommend the solution to disinfect in the following table.
Glutaraldehyde-based disinfectant:
Chemical Name Reagent Name Step
Glutaraldehyde (2,4 %) Cidex Glutaraldehyde disinfectant Please refer to the instructions of the
solution for details.
Non-glutaraldehyde-based disinfectant:
Chemical Name Reagent Name Step
Phthalaldehyde solution (0,55 %) Cidex OPA Please refer to the instructions of the
solution for details.
Please follow the instructions about disinfectant concentration and disinfection method, as well as the precautions
about disinfectants provided by disinfectant provider. But do not rinse or soak the probe connector or close to
connector cable.
The soaking time of probe in the disinfectant is limited to the minimum time recommended by disinfectant
manufacturer (e. g., Cidex OPA manufacturer recommended minimum 12 minutes).