using sterile compressed air.
Functional
Testing,
Maintenance
Visual inspection for cleanliness of the
instruments and reassembling. Functional testing
according to instructions of use. If necessary,
perform reprocessing process again until
instrument is visibly clean.
Before packaging and autoclaving, make sure
that the components have been maintained acc.
to manufacturerās instruction.
Only the contra angle needs to be lubricated.
1. Before autoclaving, the contra angle must be
lubricated.
2. Attaching the spray nozzle to oil can and
contra angle, press the oil can button more
than 3 seconds, till all the black oil flow out
from the head of the contra angle.
Pack the instruments in an appropriate
packaging material for sterilization.
1. Check the validity period of pouch given by the
manufacturer to determine the shelf life.
2. Use pouches which resist to a temperature up
to 141ā and in accordance with EN ISO 11607.
Sterilization of instruments by applying a
fractionated pre-vacuum steam sterilization
process (according to EN 285/EN 13060/EN ISO
17665) under consideration of the respective
country requirements.