11 Technical Data
Page
45
/
52
Electrostatic
discharge
(ESD) IEC
61000-4-2
+/- 8 kV
contact
+/- 2 kV, +/- 4
kV, +/- 8 kV,
+/- 15 kV air
+/- 8 kV
contact
+/- 2 kV, +/- 4
kV, +/- 8 kV,
+/- 15 kV air
Floors should
be wood,
concrete or
ceramic tile. If
floors are
covered with
synthetic
material, the
relative
humidity
should be at
least 30 %.
Electrical fast
transients/b-
ursts IEC
61000-4-4
±2kV
100kHz
repetition
frequency
±2kV
100kHz
repetition
frequency
Mains power
quality should
be that of a
typical
commercial or
hospital
environment.
Surge IEC
61000-4-5
Line to line:
±0.5kV, ±1kV
Line to earth:
±0.5kV, ±1kV,
±2kV
Line to line:
±0.5kV, ±1kV
Line to earth:
±0.5kV, ±1kV,
±2kV
Mains power
quality should
be that of a
typical
commercial or
hospital
environment.