agents to the components.
⚫ Silver tubes are single use only and shall be
disposed of according to the local environmental
protection laws and regulation.
Preparation
before
cleaning
Remove and disconnect the Cartridge nut before
cleaning. Refer to "Chapter 4-Installing the Fast-Fill"
of this manual for disassembly instructions.
Observe suitable personal protective measures.
The following Step 3 to Step 5 are operated in a washer-disinfector:
⚫ Use only approved washer-disinfectors according to EN ISO 15883,
maintain and calibrate it regularly.
⚫ Follow instructions and observe concentrations given by the manufacturer
(see general recommendations).
⚫ Sufficient rinsing step should be available in purified water (max 10
germs/ml and max 0.25 endotoxin units/ml)
⚫ Avoid any contact between the cartridge nut and any instrument, kit,
support or container.
⚫ Make sure the components are dry before moving to the #6 step.
Carefully put the components (Cartridge nut and
Bending Tool) into the washer-disinfector and set the
parameters as follows:
- Pre-cleaning: water temperature <30℃, 2 min;
- Cleaning: water temperature 45℃, 5 min; use an
enzyme detergent solution (aldehyde free
solution) which is suitable to be used with washer-
disinfector, and use in accordance with the IFU of
the detergent solution manufacturer;
- Rinsing: water temperature 45℃, 1 min (rinsing
twice).
Thermal disinfection at least 5 min at 90℃/194°F,
make sure A0 value≥3000.
Rinse with purified water at 70°C for 1 minute.
Inspect components and sort out those with defects.
Dirty components must be cleaned and disinfected