11.EMC Tables
Guidance and manufacturer’s declaration – electromagnetic emissions
The Fast-Fill is intended for use in the electromagnetic environment specified below. The
customer or the user of the Fast-Fill should assure that it is used in such an environment.
Electromagnetic environment - guidance
The Fast-Fill uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The Fast-Fill is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity
The Fast-Fill is intended for use in the electromagnetic environment specified below. The
customer or the user of the Fast-Fill should assure that it is used in such an environment.
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.