8.3.4 Disinfection components
Disinfection components
Fast-Fill Handpiece
Charge Base
Adapter
Handpiece Base
Wipe all the surfaces with a cloth lightly moistened with Ethanol for Disinfection
(Ethanol 70 to 80 vol%) at least 2min, repeat for 5 times.
NOTE
Do not use anything except Ethanol for Disinfection (Ethanol 70 to 80
vol%).
Do not use too much ethanol as it’s going into machine and damage the
components inside.
Do not spray any liquid directly on the handpiece. Do not allow any moisture
to get into the handpiece.