11 EMC Tables
Page 29 / 35
11 EMC Tables
Guidance and manufacturer’s declaration – electromagnetic emissions
The HyperLight is intended for use in the electromagnetic environment specified below.
The customer or the user of the HyperLight should assure that it is used in such an
environment.
Electromagnetic environment - guidance
The HyperLight uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The HyperLight is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity
The HyperLight is intended for use in the electromagnetic environment specified below.
The customer or the user of the HyperLight should assure that it is used in such an
environment.
Electromagnetic
environment - guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+/-8 kV contact
+/-15kV air
+/-2, 4, 6 & 8kV
contact
+/-2, 4, 8, & 15 kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are synthetic, the r/h
should be at least 30%
Electrical fast
Transient/burst
IEC 61000-4-4
+/-2 kV for
power supply
lines
+/-1 kV for input/
output lines
+/-0.5, 1 & 2 kV
for power
supply lines
+/-0.5 & 1 kV for
input/output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.