5 Product function and use
Page 19 / 35
5.4 Advanced settings
5.4.1"Cleaning" mode
It is recommended to flush the liquid
circuit of the device after scaling at
least once a day. The "Cleaning"
mode allows for cleaning the liquid
circuit in order to reduce the
accumulation of crystals and bacteria
in the liquid circuit.
Steps:
1) Install the water hose to the water
inlet correctly.
2) Press and hold and at
the same time for about 2
seconds. At this time, the buzzer
beeps once and enters the
“Cleaning” mode. At this time, the
LED digital tube on the panel
alternately displays “CL” and
cleaning time. The default
cleaning time is 30 Seconds,
press "+" or "-" key to adjust. The
adjustment range is 10 ~ 60
seconds.
3) Step on the foot pedal to start
cleaning the liquid circuit. At this
time, the foot pedal can be
released.
4) After the cleaning countdown, the
device will automatically stop and
exit the "Cleaning" mode. During
the cleaning process, step on the
foot pedal again or press the " "
key to stop cleaning and exit the
"Cleaning" mode.
5.4.2 Handpiece LED light
delay adjustment
The handpiece LED light will be lit
during the operation. The device will
stop running after the foot pedal is
released. The handpiece LED light will
be delayed for a certain time before
extinguishing. The default delay time
is 10 seconds, and the delay time can
be adjusted as needed.
Steps:
Press and hold the key and the
key at the same time for about 2
seconds, at this time the, the buzzer
beeps once and enter the “Cleaning”
mode, and then press the key to
enter the handpiece LED light
adjustment state, at this time, the LED
digital tube display the delay time. The
default is 10 seconds, you can press
the "+" or "-" key to proceed
adjustment. The adjustment range is
10 ~ 20 seconds. Press the "G" button
or no operation for about 5 seconds to
exit the setting of the handpiece LED
light delay.