EasyManua.ls Logo

Eighteeth UltraMint - 6.4 Disinfection components

Eighteeth UltraMint
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Cleaning, Disinfection and Sterilization
Page 25 / 35
6.4 Disinfection components
Disinfection components
Main unit
Power adapter
Foot pedal
Handpiece support
External water supply hose
Wipe all the surfaces with a cloth lightly moistened with Ethanol for
Disinfection (Ethanol 70 to 80 vol%) at least 2min, repeat for 5 times.
NOTE
z Do not use anything except Ethanol for Disinfection (Ethanol 70 to 80
vol%).
z Do not use too much ethanol as it’s going into machine and damage the
components inside.
z Do not spray any liquid directly on the machine. Do not allow any moisture
to get into the machine.