3
diagnoscunit
InstruconGuidelines
CauonsandWarnings
DANGER: Not for use in the presence of ammable anesthecs.
WARNING: The diagnosc unit should NOT be used on a paent with a
pacemaker.
WARNING: Do not plug any connectors or pins on the le clip or probes
into any external power source, as it may cause a severe safety hazard
to the paent.
WARNING: Only use the specied SybronEndo AC adaptor, adaptor
plugs and SybronEndo baery pack replacement.
WARNING: Use of other accessories which are not authorized for use
in connecon with this device may cause malfuncon and compromise
paent safety.
WARNING: No modicaon of this equipment is allowed.
CAUTION: This device has been invesgated with regard to safety from
electrical shock and re hazard as well as electromagnec compability
(EMC). The device has not been invesgated for other physiological
eects. Please contact Sybron if you have any further quesons
regarding electrical safety or EMC.
CAUTION: For use by qualied and trained personnel only.
CAUTION: This device to be used in conjuncon with other diagnosc
procedures and not relied on exclusively.
CAUTION: Do not autoclave the Satellite
CAUTION: This device has been tested and found to comply with EMC
limits for the Medical Device Direcve 93/42/EEC (EN 60601-1-2). These
limits are designed to provide reasonable protecon against harmful
interference in a typical medical installaon. The device generates radio
frequency energy and, if not installed and used in accordance with
these instrucons, may cause harmful interference to other devices in
the vicinity. However, there is no guarantee that interference will occur
in a parcular installaon. If this device does cause harmful interference
with other devices, which can be determined by turning the device o
and on, the user is encouraged to try to correct the interference by one
or more of the following measures:
Reorient or relocate the receiving device
Increase the separaon between devices
Connect the device into an outlet on a circuit dierent from that to
which the other device(s) is connected.
Consult the manufacturer for help
CAUTION: To reduce the risk of electrical shock do not remove cover.
Refer servicing to qualied service personnel.
CAUTION: U.S. Federal Law restricts this device to sale by or on the
order of a health care professional.
CAUTION: Remove mains plug from wall socket to isolate unit from
mains power.
CAUTION: Posion device such that it can be easily unplugged from the
wall.
Parts List
diagnosc unit 973-0300
diagnosc unit (No Satellite) 973-0320
AC Adaptor 973-0302
Satellite Cord with Satellite 973-0303
Baery Pack 973-0305
Paent Lead Cord 973-0311
Bifurcated Probe 973-0307
File Clip 973-0310
Vitality Probe 973-0308
Crown Test Tip 973-0309
OponalAccessories
Satellite Cord w/o Satellite 973-0304
Mounng Bracket 973-0306
Symbols
Adjustment of vitality rate, audio volume, LED brightness
Mode switch between Apex mode and Vitality mode
Setup
On
Corresponds to MDD 93/42 EEC
Including EN 60601-1 and EN 60601-1-2
European Representave
CSA mark with “C/US” indicator for cered products
Aenon! Consult manual before use
Consult accompanying documents
Type BF Paent Applied Part
Class II Medical Electrical Equipment
Do not autoclave
Part Number
Serial Number
Manufacturer
Manufacturing Date
This way up
Fragile, handle with care
Keep away from rain
+45°C
-10°C
Temperature limits for storage and transportaon
Humidity Limitaon
Pressure Limitaon
Do not throw into trash
ProductFeatures
The Diagnosc Unit combines a Vitality Scanner and an Apex Locator in
one unit.
ApexLocator(AL)Mode
The Apex Locator calculates the distance from the p of your
endodonc le to the foramen by measuring changes in impedance
between two electrodes. The rst electrode is the lip clip. The second is
the probe or le clip which makes contact with a le or plugger that has
been inserted into the root canal.
The units advanced technology allows it to give accurate readings in the
presence of sodium hypochlorite, anesthec, alcohol, EDTA, blood, pus,
etc., as well as in the presence of necroc ssue.
VitalityScanner(VS)Mode
The Vitality Scanner’s pulsang smulus allows for determinaon of
pulpal vitality. The smulus strength automacally increases from a low
value to prevent discomfort and is electronically stabilized to provide
consistent results. Some paents may experience slight discomfort
during the test.
Aer a tooth is tested, the display retains its response level to facilitate
recording of results. The unit automacally resets to aid rapid tesng of
mulple teeth.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Elements SybronEndo and is the answer not in the manual?

Elements SybronEndo Specifications

General IconGeneral
BrandElements
ModelSybronEndo
CategoryDental equipment
LanguageEnglish