Manual 4 Point
Manual 4 PointFigure 4-3:
a. Position the laser and sensor at a starting angle (ideally at the first
cardinal position, 0°).
The laser and sensor angles should be within 2° of each other and
within 3° of the active cardinal position (0°, 90°, 180°, and 270°), if
possible. When this happens, the background of the box displayed
below the machine is green. If the laser and sensor angles are not
within 2° of each other and 3° of the active cardinal position, the box
background is white.
b.
Rotate the laser fixtures a full revolution stopping at each cardinal
position (0°, 90°, 180°, and 270°). Press F3 Accept Readings after
stopping at each point.
Note
While it is always desirable to measure at all four cardinal positions, it is
possible to estimate an alignment solution with only three
measurements.
Quick Start Guide December 2018
26 AMS 2140 Advanced Laser Alignment Application