Procedure
1. If desired, re-orient the transmitter on the mounting bracket.
a. Remove the junction end-cap from the junction housing.
b. Loosen each of the four cap screws (4 mm).
c. Rotate the bracket so that the transmitter is oriented as desired.
d. Tighten the cap screws, torquing to 30 to 38 in-lbs (3 to 4 N-m).
e. Replace the junction end-cap.
Components of 4-wire remote mount transmitter (aluminum
housing)
Figure 3-3:
B
D
C
A
A. Transmitter
B. Mounting bracket
C. Cap screws
D. End-cap
Mounting and sensor wiring for 4-wire remote installations
Installation Manual 19