• Micro Motion recommends the use of 5/16-18 (8 mm–1.25) fasteners that can
withstand the process environment. Micro Motion does not supply bolts or nuts as
part of the standard offering (general purpose bolts and nuts are available as an
option).
• Ensure that the surface is flat and rigid, does not vibrate, or move excessively.
• Confirm that you have the necessary tools, and the mounting kit shipped with the
transmitter.
For mounting the remote core processor to an instrument pole:
• Use two 5/16-inch U-bolts for 2-inch pipe, and four matching nuts, that can
withstand the process environment. Micro Motion does not supply U-bolts or nuts
(appropriate bolts and nuts are available as an option).
• Ensure the instrument pole extends at least 12 inches (305 mm) from a rigid base,
and is no more than 2 inches (50.8 mm) in diameter.
Procedure
1. If desired, reorient the core processor housing on the bracket.
a. Loosen each of the four cap screws (4 mm).
b. Rotate the bracket so that the core processor is oriented as desired.
c. Tighten the cap screws, torquing to 30 to 38 in-lbs (3 to 4 N-m).
Components of a remote core processorFigure 5-5:
A. Mounting bracket
B. Cap screws
2. Attach the mounting bracket to an instrument pole or wall.
5.3 Prepare the 4-wire cable
Important
For user-supplied cable glands, the gland must be capable of terminating the drain wires.
Mounting and sensor wiring for remote core processor with remote sensor installations