User-supplied 4-wire cable must meet the following requirements:
• Twisted pair construction.
• Applicable hazardous area requirements, if the core processor is installed in a
hazardous area.
• Wire gauge appropriate for the cable length between the core processor and the
transmitter.
Wire gaugeTable 3-1:
Wire gauge Maximum cable length
VDC 22 AWG (0.35 mm
2
) 300 ft (90 m)
VDC 20 AWG (0.5 mm
2
) 500 ft (150 m)
VDC 18 AWG (0.8 mm
2
) 1000 ft (300 m)
RS-485 22 AWG (0.35 mm
2
) or larger 1000 ft (300 m)
3.3 Wire the transmitter to the sensor
1. Connect the cable to the sensor-mounted core processor as described in the sensor
documentation.
2. Feed the wires from the sensor through the conduit opening on the transmitter.
3. Connect wires to the appropriate terminals on the mating connector.
Tip
You may find it easier to unplug the mating connector to connect the wires. If you do so,
remember to firmly reseat the mating connector and tighten the mating connector screws so
that the mating connector cannot accidentally come loose.
Mounting and sensor wiring for 4-wire remote installations
Installation Manual 23