Sensor and transmitter terminal designationsTable 4-5:
Wire color Sensor terminal Transmitter terminal Function
Black No connection 0 Drain wires
Brown 1 1 Drive +
Red 2 2 Drive –
Orange 3 3 Temperature –
Yellow 4 4 Temperature return
Green 5 5 Left pickoff +
Blue 6 6 Right pickoff +
Violet 7 7 Temperature +
Gray 8 8 Right pickoff –
White 9 9 Left pickoff –
d. Tighten the screws to hold the wire in place.
e. Ensure integrity of gaskets, grease all O-rings, then replace the junction box and
transmitter housing covers and tighten all screws, as required.
4.3.1 Sensor and transmitter terminals
All ELITE, H-Series, and T-Series sensor, and 2005 or newer F-Series sensor
terminals
Figure 4-8:
A. Violet
B. Yellow
C. Orange
D. Brown
E. White
F. Green
G. Red
H. Gray
I. Blue
Mounting and sensor wiring for 9-wire remote installations
Installation Manual 37