3. Feed the wires from the remote core processor through the conduit opening.
4. Connect wires to the appropriate terminals on the mating connector.
Tip
You may find it easier to unplug the mating connector to connect the wires. If you do so,
remember to firmly reseat the mating connector and tighten the mating connector screws so
that the mating connector cannot accidentally come loose.
Wiring path for transmitters with aluminum housingFigure 5-9:
A
C
VDC+
VDC
RS-485A
RS-485B
B
A. 4-wire cable
B. Transmitter conduit opening
C. Mating connector
Mounting and sensor wiring for remote core processor with remote sensor installations
Installation Manual 57