Wire the frequency output (internally powered (Channel B)
Procedure
Wire to the appropriate output terminal and pins.
Figure 3-14: Frequency output wiring (internally powered)
A. Frequency output
B. See
Figure 3-15 for output amplitude versus load resistance
C. Counter
Figure 3-15: Internally powered frequency output: output amplitude versus load
resistance [24 VDC (Nom) open circuit]
A. Output amplitude (V) where V= 22mA x Load Resistor in Linear Range
B. Load resistor (Ω)
C.
Linear Range