Byte Access Name Type Notes
4–7 Get/Set External Pressure Float
8 Get/Set Start/Stop Totals Byte 0: Stop
1: Start
9 Get/Set Reset All Process Totals Byte 0: No action
1: Reset
10 Get/Set Reset All Inventory Totals Byte 0: No action
1: Reset
11 Get/Set Start Zero Byte 0: Abort or no action
1: Start
12–13 Get/Set Start Smart Meter
Verification
Word See Table B-20
B.3.3 Input parameters (explicit data) for standard
configuration
To update any of these attribute values, the associated trigger byte must be toggled
before reading the attribute value (executing the Get service). See Output parameters
(explicit data) for standard configuration for more information on trigger bytes.
Class B0h, Instance 01h
Attribute
Access Name Type Trigger Byte
WriteAttribute
Description
01h Get Mass flow rate unit U16 0Ch See Table B-7
02h Get Density unit U16 0Ch See Table B-13
03h Get Temperature unit U16 0Ch See Table B-14
04h Get Volume flow rate unit U16 0Ch See Table B-9
05h Get Pressure unit U16 0Ch See Table B-15
06h Get Mass total/inventory unit U16 0Ch See Table B-8
07h Get Volume total/inventory unit U16 0Ch See Table B-10
08h Get Zero time U16 0Dh Seconds
09h Get Standard deviation of auto
zero
Float 0Eh
0Ah Get Present flow signal offset at
zero flow
Float 0Eh
0Bh Get Failed Zero Calibration Value Float 0Eh
User Guide Device profile
MMI-20019808 January 2019
User Guide 45

Other manuals for Emerson Micro Motion

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion and is the answer not in the manual?

Emerson Micro Motion Specifications

General IconGeneral
CategoryMeasuring Instruments
ManufacturerEmerson
BrandMicro Motion
TypeCoriolis Flow Meter
Output4-20 mA, HART, Foundation Fieldbus, Profibus, Modbus
Power Supply24 V DC
MaterialStainless Steel, Hastelloy
Process ConnectionFlanged, Threaded
CertificationsATEX, IECEx, CSA, SIL
Pressure RatingVaries by model; up to 10, 000 psi (690 bar)
Temperature RangeVaries by model; -400 to 750°F (-240 to 400°C)

Related product manuals