1.3 Specifications
1. Requires XPS transmitter, P/N 6A00358G03
Note
All static performance characteristics are with operating variables constant. Specifications subject to
change without notice.
1.3.1 Net O
2
range
0-1% to 0-40% O
2
, fully field selectable
1.3.2 Combustibles
0-1,000 ppm to 0-5%, fully field selectable
1.3.3 Accuracy
Oxygen
±0.75% of reading or 0.5% O
2
(whichever is greater)
Combustibles
±2% range
1.3.4
System response to test gas
Oxygen
10 sec T90
Combustibles
25 sec T90
1.3.5
Temperature limits
Process
0 to 1427 °C (32 to 2600 °F)