ENGLISH
FB-618/AU rev.E – ed.2019/08
14/48
2.4. Water supply and PIEZON
®
bottle
Without Bottle:
PIEZON
®
& AIRFLO W
®
use external water supply.
With Bottle connected:
AIRFLOW
®
uses external water supply.
PIEZON
®
uses bottle liquid supply.
The CLIP+CLEAN shall be
previously cleaned and disinfected
before use.
Non-disinfected CLIP+CLEAN may
contaminate the device.
Place the CLIP+CLEAN into the
device’s
bottle receptacle for dust
protection.
Connect the PIEZON
®
bottle
Only use the PIEZON
®
bottle EG-111
(transparent) for disinfectant solutions.
The PIEZON
®
bottle is compatible with the following solutions:
Avoid unwanted chemical reactions or ingestion of residues of solution. Always rinse the liquid
circuit with drinking water before using a different solution or undergo a regular treatment requiring
water only.
For treatment application use the percentage of solution as per discretion and per local country
regulations and recommendations.
Solution
Concentration
Sodium Hypochlorite
Up to 5%
Chlorexidine
Up to 1%
Carbanilides
Up to 0.05%
Povidone iodine
Up to 12%
Quaternary ammonium compound
Up to 0.7%
Cetylpyridin chloride
Up to 1%
Citric acid
Up to 5%
Physiological water (NaCl %)
Up to 0.9%
Ethanol
Up to 5%
Isopropanol
Up to 5%
Do not sterilize the PIEZON
®
bottle and its nozzle cap by steaming or dry thermal reprocessing.
Use only ambient temperature active disinfectant and cleaning agents.