6.6. Yearly maintenance & repair
This device must only be maintained and/or repaired by EMS and by
authorized EMS repair centers.
A yearly preventive maintenance
or 2000 hours usage maintenance
(LED solid orange), whichever comes first, is required as means of
safety and performance guarantee for both the patient and the user.
Qualified service repair may also be required anytime persistent malfunctioning is detected
by the user and/or reported by the device diagnostic.
When returning the device for service, it is recommended that you ship the
device with its pedal, powder chamber, bottle and cords in its original
packaging for optimal protection against damage during transportation.
Provide the contact details of your EMS dealer for a quicker service process (see § 6.9).
6.7. Pairing a new pedal
1. Remove one battery from the pedal (no need to remove both).
2. Place the two handpieces in their holders.
3. Turn the machine OFF, wait 10 seconds, then turn it ON again.
4. Press + first, then also press simultaneously.
A sonar sound will start playing (if not, repeat step 4).
Respect the order and the three-finger sequence (see figure below place fingers
in the groove below the numbers).
5. While the sonar sound plays, replace the lithium batteries into the
wireless pedal.
6.
Within a short time (less than 15 seconds), the pairing will be
complete, the white LEDs will blink for a while and the device is
then ready for use.
-
If the process takes longer than 1 minute, it means the pairing has failed and the device will
automatically exit the mode. (No more sonar sound and no blinking at exit).
In case of this process failure, redo it from the beginning.

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz