6. MAINTENANCE & TROUBLESHOOTING
6.1. AIRFLOW
®
Handpiece powder unclogging
In case of a clogged handpiece and before the reprocessing of AIRFLOW
®
and PERIOFLOW
®
Handpieces.
Easy Clean
Provided in your
®
Application box
Rinse through central lumina in
the normal direction of flow (no back
rinsing) using Easy Clean with a
disposable syringe filled with more
than 2 ml of drinking water
The Easy Clean tool can be thermally disinfected and also sterilized at up to 135°C in the autoclave.
6.2. AIRFLOW
®
handpiece leakage
In case of leakage at the AIRFLOW
®
handpiece
connection with the AIRFLOW
®
cord, replace the o-rings
of the cord with the spare provided in the EL-
located in the
®
application box.
6.3. PIEZON
®
light guide check & replace
The light guide loses its transparence after undergoing repeated reprocessing cycles. Check the
transparency of the light guide every month and do the following:
1. Remove the tip and unscrew the handpiece nose cap by hand.
2. Take off the light guide and inspect it.
3. Place in a new light guide AB-340 (provided in the
®
application).
4. Screw the nose cap on again, by hand only.