1.5. Contraindications
Treatments contraindications:
Suggestion for alternatives:
®
and
®
are contraindicated
with
Patients with severe or unstable upper
respiratory tract infections, chronic
bronchitis/asthma
1
.
PIEZON
®
®
is contraindicated with
Pregnant and breastfeeding
patients
AIRFLOW
®
and PIEZON
®
PS
®
is contraindicated with
Patients with severe inflammation and/or
osteonecrosis.
AIRFLOW
®
PLUS
and PIEZON
®
PS
®
is contraindicated with
Patients with a cardiac pacemaker,
defibrillators and any implantable electronic
device.
AIRFLOW
®
PLUS
The decision to use AIRFLOW
®
and/or PERIOFLOW
®
on contagious patients or on patients with
risk of infection, has to be taken by the dentist/medical doctor on an individual basis following
practitioner protection level, patient risk assessement and specific country regulations.
On patients under Bisphosphonate therapy, the decision to use AIRFLOW
®
and/or
PERIOFLOW
®
has to be taken by the dentist/medical doctor depending on the oral health of the
patient.
AIRFLOW
®
powders contraindications:
Suggestion for alternatives:
CLASSIC powder
is contraindicated with
Low-salt diet patients.
AIRFLOW
®
PLUS
Flavored CLASSIC
powder
is contraindicated with
Patients allergic to flavor aroma.
AIRFLOW
®
PLUS/PERIO
and CLASSIC NEUTRAL
PLUS
powder
is contraindicated with
Patients allergic to Chlorhexidine.
AIRFLOW
®
PERIO
PERIO
&
SOFT
powder
is contraindicated with
Patients allergic to Glycine
(Glycocoll).
AIRFLOW
®
PLUS
1.6. Compatibility
This device is compatible with the following accessories:
®
Powders
PLUS powders: DV-082, DV-086 series
CLASSIC powders: DV-048 series
PERIO and SOFT powders: DV-070, DV-071 series
®
Handpiece
EL-308
®
Handpiece
EL-354
®
Handpieces
EN-060, EN-061
®
Scaling and Periodontal instruments
PS, A, B, C, P, PSR, PSL, PL1, PL2, PL3, PL4, PL5,
HPL3, DPL3, PI
®
Endodontics instruments
RT1, RT2, D, H, ESI, Files ISO 15, 20, 25, 30, 35,
Endochuck 180°, 120° and 90°, RT3, RE2, BERUTTI
®
Conservative measures instruments
E, F, G, SP
®
Cavity preparation instruments
SM, PF, SD, VE, SB, SBD, SBM
1
Related to possible powder inhalation during AIRFLOW
®
treatment.

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz