EasyManuals Logo

EMS AIRFLOW Prophylaxis Master User Manual

EMS AIRFLOW Prophylaxis Master
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
ENGLISH
FB-618/AU rev.E – ed.2019/08
26/48
5.3. Reprocessing of EMS parts
EMS recommends the use of cleaning, disinfection, packaging for sterilization and sterilization
procedures accordingly with ISO 17664.
Always report adverse events related to device reprocessing directly to EMS.
Reusable products must be cleaned disinfected and, if applicable, sterilized prior to first use. Do
not reprocess the products over the allowed number of sterilization cycles, but replace: refer to the
“Service life” section of the “Technical Descriptionchapter.
Concentrations and contact times specified by the manufacturer of the cleaning and disinfection
agent must be followed.
Remember that sterilization cannot be achieved unless the elements of the assembly are
cleaned and disinfected first.
If there is anything in the following instructions that is not clear or seem to be inadequate, do not
hesitate to contact/inform EMS.
The following instructions have been validated as being capable of preparing for re-use the EMS
medical devices and parts listed in the “Intended Use and Compatibility” chapter. It remains the
responsibility of the user to ensure that the processing as actually performed using equipment,
materials and personnel in the processing facility achieve the desired result. This requires validation
and routine monitoring of the process. Likewise any deviation by the processor from the
instructions provided should be properly evaluated for effectiveness and potential adverse
consequences.
The user shall also observe any applicable legal requirements in their country as well as the hygiene
regulations of the hospital or clinic. This applies especially with regard to the additional
requirements for the inactivation of prions.

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz

Related product manuals