Preparation
Manual pre-cleaning is required:
Immediately after use, rinse the lumen(s) line of the handpiece/instrument with water for 20
seconds. Coarse soiling must be removed immediately after application.
For AIRFLOW
®
and PERIOFLOW
®
: always carry out handpiece powder unclogging and check
for both lumens (water and powder) clearage before proceeding.
For PIEZON
®
:
Remove any installed instrument.
Remove the nose cap and separate aside the light guide and the O-ring.
For Endochuck file holder:
Remove any installed file.
Remove the screw and separate aside the small O-ring.
For any Instrument mounted on the CombiTorque:
Separate aside the Instrument and the CombiTorque.
Safely transport to the reprocessing area to avoid any damage to the parts and contamination to
the environment and to the people involved in the reprocessing process.
Cleaning shall need to be performed within 1 hour from the use.

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz

Related product manuals