Cleaning
Any part can be cleaned manually or automatically by washer or disinfector. EMS recommends the
use of an ISO 15883 compliant automatic washer-disinfector (WD) for an optimal effectiveness and
part service life.
DO NOT use any Ultrasound Bath cleaning procedure with the PIEZON
®
, AIRFLOW
®
and
PERIOFLOW
®
handpieces: it may destroy the products.
Manual Cleaning (without ultrasound bath)
Automated Cleaning
The following validated process can be used
with any EMS part:
Remove away any surface soiling (gross
contamination) on the product with a
wetted cloth and tap water.
Brush the devices in a cleaning solution
of 0.5% neodisher MediClean (Dr.
Weigert Hamburg) in deionized water at
40°C with a suitable soft bristled brush
(Medisafe MED100.33) until all visible
residues have been removed.
For parts having lumens, flush all lumens
with a spray gun (water jet gun, with a
static water pressure of 2 bar) with cold
tap water for 15 seconds.
Immerse the part in 0.5% solution of
neodisher MediClean with deionized
water for 15 minutes at 40°C. Taken
care that all lumens are filled with
cleaning solution (use a syringe if
needed). All surfaces must be
moistened.
Rinse all lumens flushing with a spray
gun (water jet gun, with a static water
pressure of 2 bar) with cold deionized
water for 15 seconds, and further rinse
the whole part under cold running
deionized water for 10 seconds.
Using an air pistol (compressed air) fully
dry the lumen and the whole part until no
more residues of water are present
(visible or detectable).
The washer-disinfector
shall comply with
ISO 15883 standard,
must have suitable
baskets to hold small fragile products and must
have rinsing connections
with diameter of
approximately 16mm
for the attachment to
product lumen.
Correctly place the instrument into a suitable
rack
, connect all lumens to the rinsing
connectors and start the automated cleaning
.
The following validated (For example on Miele
Professional G 7836 CD having Miele Rack
E429) automated cleaning process
can be
used:
2 minutes pre-washing with cold water.
Drain.
5 minutes cleaning with tap water and 0.5%
detergent of neodisher MediClean Dental
(Dr. Weigert, Hamburg) at 55°C.
Drain.
3 minutes rinsing and neutralization cold
with deionized water.
Drain.
2 min final rinse with cold deionized water.
Drain.
Also instructions of the manufacturer of the
washer disinfector shall be followed.

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz