Manual Cleaning with Ultrasound Bath
DO NOT USE WITH HANDPIECE BODY
The following validated process
is intended
ONLY for the EMS PIEZON Instruments and file
holders:
Remove away any surface soiling (gross
contamination) on the product with a
wetted cloth and tap water.
For part having lumens, flush all lumens
with a spray gun (water jet gun, with a
static water pressure of 2 bar) with cold
tap water for 15 seconds.
Using an appropriate sieve for
sonication, immerse the part in 0.5%
solution of neodisher MediClean with
deionized water and sonicate for 10
minutes at 40°C (Bandelin, Sonorex
1028K, 35kHz). Taken care that all
lumens are filled with cleaning solution
(use a syringe if needed). All surfaces
must be moistened.
Rinse all lumens flushing with a spray
gun (water jet gun, with a static water
pressure of 2 bar) with cold deionized
water for 15 seconds, and further rinse
the whole part under cold running
deionized water for 10 seconds.
Using an air pistol (compressed air) fully
dry the lumen and the whole part until no
more residues of water are present
(visible or detectable).
Example of correct placement of the parts
in the WD Miele Professional G 7836 CD
using the Mobile Injector Unit (Rack) Miele E429

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz

Related product manuals