ENGLISH
FB-618/AU rev.E – ed.2019/08
24/48
Between each patient
Overall cleaning and disinfection
Clean the external surface of the device
with a cloth and alcohol
Reprocess handpieces and instruments
See the specific following chapters.
Clean the unit only with an alcohol-based (ethanol,
isopropanol), colorless disinfectant.
Never use scouring powder or an abrasive sponge. It
will damage its surface.
Risk of contamination. Always disinfect the bottom
and top areas of device air connections.
End of day: Overnight cleaning
Use only EMS NIGHT CLEANER as a cleaning agent.
Other products may damage or not clean the unit, and cause patient intoxication.
Place the NIGHT
CLEANER bottle onto
the device
Set water to 10
Turn the device ON
Hold both cords over a
sink with CLIP+CLEAN
Press the pedal once,
release, and then wait 1
minute
Before placing, remove
CLIP+CLEAN from the device.
Each cleaning consumes 30ml
of NIGHT CLEANER.
Before cleaning, check
that the liquid level is above the
black flange of the bottle’s
neck.
Set both water regulators
to 10 to ensure the flow of the
cleaning agent.
Contamination prevention:
Do not make any contact
between the sink and the
cords.
CLIP+CLEAN shall be
reprocessed after each use.
The white and blue countdown
indicates remaining time.
Cleaning can be paused and
resumed by pressing the pedal
again.
Once completed, leave the
NIGHT CLEANER bottle on the
device overnight at least 12h
minimum.
1
2
3
4

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz