Wired pedal does not work
1° Disconnect and reconnect the pedal. Check the cable for damage. Restart the device.
2° If still not solved, contact EMS aftersales service.
No pressurization of the powder chamber
Check that your device is ON: at least 1 LED light should be ON.
2° Check that the AIRFLOW
®
cord system is well connected (full green mark on the lock actuator).
3° If still not solved, contact EMS aftersales service.
Powder chamber white light is BLINKING at pressurization attempt
Either the air line is not connected or there is not enough air pressure.
1° Check the air line for no kinking and check the air compressor unit.
2° Check air filter for cleanliness and replace if dirty.
3° If still not solved, contact EMS aftersales service.
Powder chamber white light is BLINKING at depressurization
The handpiece could be clogged. Unclog with Easy Clean (see paragraph below).
2° AIRFLOW
®
cord could be clogged. Dismount and clean the airflow cord extremity.
3° If still not solved, contact EMS aftersales service.
Powder sprays out of chamber at depressurization
1° Powder chamber is filled beyond the maximum level marked.
2° Remove the powder exceeding the MAX sign on the bottle.
Powder leaks under the AIRFLOW
®
Handpiece cord system
The AIRFLOW
®
pinching element might be worn out or the air interface is dirty and is leaking powder.
1° Disconnect the cord, clean the air jack and connect again. If problem persists, go to Step 2.
2° Replace your AIRFLOW
®
Handpiece cord with a new one.
3° If still not solved, contact EMS aftersales service.
Powder chamber is leaking
Clean the chamber with a wet cloth, in particular the cap and the bottom o-rings. Also clean the
connecting elements on the device.
2° If still not solved, replace the powder chamber with a new one.
White LED PIEZON
®
is not working
1° Clean and dry the handpiece connection and try again.
2° Your PIEZON
®
Handpiece LED might have been switched off by activity time-outs:
- after 10 minutes of continuous operation,
- or after 20 seconds of inactivity off the holder.
In both cases, put back the handpiece into the holder, wait 1 minute and try again.
3° If still not solved, contact EMS aftersales service.
Insufficient lighting
1° Check the light guide and replace if necessary.
2° If the light is still weak, replace the handpiece.
Damaged light guide
Replace the light guide.
Low or no mechanical power delivered by PIEZON
®
or vibration perceived
1° Make sure that the PIEZON
®
Instrument (tip) is correctly screwed on (use CombiTorque
®
tool).
2° Check the wear of the instrument (tip) and replace it if necessary.
Clean and dry the handpiece and the cord system electric connections.
4° Replace the PIEZON
®
Handpiece first.
5° Replace the PIEZON
®
Handpiece cord.
6° If still not solved, contact EMS aftersales service.

Other manuals for EMS AIRFLOW Prophylaxis Master

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW Prophylaxis Master and is the answer not in the manual?

EMS AIRFLOW Prophylaxis Master Specifications

General IconGeneral
Water Flow RateAdjustable
Powder Flow RateAdjustable
TechnologyAIRFLOW, PERIOFLOW, PIEZON
Water SupplyIntegrated water tank
ControlFoot control
HandpiecesAIRFLOW, PERIOFLOW, PIEZON
Water tank capacityApprox. 800 ml
Water pressure1 - 5 bar (100 - 500 kPa)
Operating modeContinuous
TypeAir polishing system
Powder TypeErythritol, Glycine
Nozzle TypesStandard, Perio
Power Supply100-240 V, 50/60 Hz