FB-618/AU – rev.E – ed.2019/08
Wired pedal does not work
1° Disconnect and reconnect the pedal. Check the cable for damage. Restart the device.
2° If still not solved, contact EMS aftersales service.
No pressurization of the powder chamber
1° Check that your device is ON: at least 1 LED light should be ON.
2° Check that the AIRFLOW
®
cord system is well connected (full green mark on the lock actuator).
3° If still not solved, contact EMS aftersales service.
Powder chamber white light is BLINKING at pressurization attempt
Either the air line is not connected or there is not enough air pressure.
1° Check the air line for no kinking and check the air compressor unit.
2° Check air filter for cleanliness and replace if dirty.
3° If still not solved, contact EMS aftersales service.
Powder chamber white light is BLINKING at depressurization
1° The handpiece could be clogged. Unclog with Easy Clean (see paragraph below).
2° AIRFLOW
®
cord could be clogged. Dismount and clean the airflow cord extremity.
3° If still not solved, contact EMS aftersales service.
Powder sprays out of chamber at depressurization
1° Powder chamber is filled beyond the maximum level marked.
2° Remove the powder exceeding the MAX sign on the bottle.
Powder leaks under the AIRFLOW
®
Handpiece cord system
The AIRFLOW
®
pinching element might be worn out or the air interface is dirty and is leaking powder.
1° Disconnect the cord, clean the air jack and connect again. If problem persists, go to Step 2.
2° Replace your AIRFLOW
®
Handpiece cord with a new one.
3° If still not solved, contact EMS aftersales service.
Powder chamber is leaking
1° Clean the chamber with a wet cloth, in particular the cap and the bottom o-rings. Also clean the
connecting elements on the device.
2° If still not solved, replace the powder chamber with a new one.
White LED PIEZON
®
is not working
1° Clean and dry the handpiece connection and try again.
2° Your PIEZON
®
Handpiece LED might have been switched off by activity time-outs:
- after 10 minutes of continuous operation,
- or after 20 seconds of inactivity off the holder.
In both cases, put back the handpiece into the holder, wait 1 minute and try again.
3° If still not solved, contact EMS aftersales service.
1° Check the light guide and replace if necessary.
2° If the light is still weak, replace the handpiece.
Replace the light guide.
Low or no mechanical power delivered by PIEZON
®
or vibration perceived
1° Make sure that the PIEZON
®
Instrument (tip) is correctly screwed on (use CombiTorque
®
tool).
2° Check the wear of the instrument (tip) and replace it if necessary.
3° Clean and dry the handpiece and the cord system electric connections.
4° Replace the PIEZON
®
Handpiece first.
5° Replace the PIEZON
®
Handpiece cord.
6° If still not solved, contact EMS aftersales service.