3.7. INSTALLING THE SINGLE-USE FLUID
MANAGEMENT SYSTEM SET (OPTIONAL) AND
REPLACEMENT POUCH
For the peristaltic pump device only
1. Suspend the two uid pouches, on the cart or on an
IV pole, at a level that is lower than the console.
Figure 28
2. Connect the uid management system input tube (A)
to the stone catcher output tube connector.
Figure 29
3. Close clamp (B) of one pouch to ll the rst pouch.
Clamp (C) stays open.
A
B
C
Figure 30
4. When the open pouch is lled, open the closed clamp
(B) rst.
5. Close the open clamp (C) (adjacent to the lled pouch).
6. The lled pouch can be exchanged for a new empty
pouch, using the Luer-lock connection.
Figure 31
3.8. CONNECTING THE STERILIZED HANDPIECE
TO THE CONSOLE
Make sure that the handpiece connector is dry be-
fore connecting it to the console.
1. To remove the protective cap from the handpiece cord,
hold the metal part of the handpiece cable connector
and push up on the cap using your thumb and index
nger.
Figure 32