5.2. PROBE INSERTION
Do not touch the probe during activation.
If a probe breaks distally, use sterile grasping
forceps to remove probe pieces from the urinary
tract.
1. Throughout the entire treatment, keep the probe tips
under endoscopic vision.
To avoid bending the probe, make sure that the
probe and the endoscope are aligned.
The probe tip should be extended 10 - 20 mm
beyond the endoscope tip.
2. Introduce and position the probe inside the endoscope.
3. The probe shall be in contact with the stone.
4. Make sure that the operation is performed with
continuous endoscopic vision.
5.3. TREATMENT SETTINGS
1. The probe is automatically recognized by the
handpiece to congure the console parameters for
each probe type.
2. The READY screen will display factory settings or the
settings used for the previous treatment.
For peristaltic pump device
Figure 53
For pinch valve device
Figure 54
3. All probe and handpiece usage information are
automatically recorded in the console (number of
uses, time of use, etc.).
4. According to the type of treatment, two pre-settings
are available:
Hard Stones Treatment,
Soft Stones Treatment.
5. You can also set each parameter manually.
Refer to the following sections:
- Custom Settings,
- Hard Stones Treatment Settings,
- Soft Stones Treatment Settings.
5.3.1. Custom Settings
1. From the STAND BY screen, press the START button.
Figure 55